Patient- and Technician-Oriented Attitudes Toward Transcranial Magnetic Stimulation Devices

Publication: The Journal of Neuropsychiatry and Clinical Neurosciences

Abstract

Four transcranial magnetic stimulation (TMS) devices are currently approved for use in treatment-resistant depression. The authors present the first data-driven study examining the patient- and technician-experience using three of these distinct devices. A retrospective survey design with both patient and technician arms was utilized. The study population included patients who received TMS for treatment-resistant depression at the Berenson Allen Center for Noninvasive Brain Stimulation for the first time between 2013 and 2016 and technicians who worked in the program from 2009 to 2017. Statistical analysis included t tests and analyses of variance to assess differences between and across the multiple groups, respectively. Patients treated with the NeuroStar device reported greater confidence that the treatment was being performed correctly compared with those treated with the Magstim device. Conversely, with regard to tolerability, patients treated with the Magstim device reported less pain in the last week and less pain on average compared with those treated with the NeuroStar device. On average, technicians reported feeling that both the Magstim and NeuroStar devices were significantly easier to use than the Brainsway Deep TMS H-Coil device. Additionally, they found the former two devices to be more reliable and better tolerated. Furthermore, the technicians reported greater confidence in the Magstim and NeuroStar devices compared with the Brainsway Deep TMS H-Coil device and indicated that they would be more likely to recommend the two former devices to other treatment centers.
Transcranial magnetic stimulation (TMS) offers a meaningful treatment option for patients with medication-refractory major depressive disorder.13 The Food and Drug Administration has approved four distinct TMS devices for treatment-resistant depression: the NeuroStar TMS Therapy System (Neuronetics, Malvern, Pa., 2008),4 the Deep TMS H-Coil System (Brainsway, Jerusalem, 2013),5 the Magstim Rapid2 Therapy System (Magstim Company, Morrisville, N.C., 2015),6 and the Magvita TMS Therapy device (MagVenture, Alpharetta, Ga., 2015).7 The first three devices have been in use in recent years as clinical treatment options at the Berenson Allen Center for Noninvasive Brain Stimulation at Beth Israel Deaconess Medical Center (Boston).
Across devices, the underlying principles of noninvasive brain stimulation are broadly overlapping. However, each machine utilizes a unique approach to achieve preferential stimulation over the dorsolateral prefrontal cortex. While the Magstim6 and Magvita8 devices both feature what is known as a “figure of 8” coil,” the NeuroStar4 machine uses a proprietary variation on this design, with some differences in shape and function. The Brainsway Deep TMS H-Coil5 device features yet another variation, utilizing an H-coil that incorporates multiple coil loops for synergistic magnetic field creation.
Little is known about how these devices compare with regard to patient and technician experience. The aim of this study was to obtain quantitative data regarding these experiences with each distinct device. We hypothesized that while no device has a demonstrated advantage in efficacy, there would be notable differences in perception of tolerability and usability across several domains of interest.

Methods

A retrospective survey design with both patient and technician arms was used. The study was approved under exempt status by the institutional review board at Beth Israel Deaconess Medical Center, and we adhered to ethical standards set forth by the institution, including that all participants provide informed consent. The study population included patients who received TMS for treatment-resistant depression at the Berenson Allen Center for Noninvasive Brain Stimulation for the first time between 2013 and 2016. One hundred fifty-seven patients were contacted, and 67 (43%) were reached and agreed to participate. All participants provided informed consent and completed a telephone survey consisting of 10 Likert-type quantitative items regarding their treatment experience. Only five patients who received treatment with the Deep TMS H-Coil System device were enrolled in the study. Given this low number of participants, patient data for this device were excluded from the comparative analysis. However, descriptive findings for all three devices analyzed (Magstim, NeuroStar, and Deep TMS H-Coil System) are summarized in Table 1. The technician arm involved a similar survey consisting of five Likert-type quantitative items pertaining to each device. This questionnaire was sent to all technicians who worked at the Berenson Allen Center between 2009 and 2016. Of the 10 individuals solicited, all completed the survey. The year 2009 was chosen as the start date because it was the first year that multiple devices were used clinically at the center.
TABLE 1. Average Patient Ratingsa
VariableNeuroStarMagstimBrainsway Deep TMS H-Coil System
Pain   
 First week2.542.182.41
 Last week1.74*1.25*1.38
Noise2.562.882.69
Would choose same device again3.613.463.47
Felt safe4.744.834.8
Anxiety (average first and last week)1.752.071.98
Confidence in device4.54*4.07*4.34
Felt tired after treatment1.742.261.92
Hamilton Depression Rating Scale % change46.944.918.5
Beck Depression Inventory % change52.343.035.3
a
Average patient ratings were on a scale from 1 to 5, ranging from “none” to “severe” and “strongly disagree” to “strongly agree.” Bold indicates statistical significance.
*
p<0.05.
Statistical analysis (performed using JMP [www.jmp.com]) included t tests and analyses of variance (ANOVAs) to assess differences between and across the multiple groups, respectively. Statistical significance was defined in these analyses as a p value <0.05, but values that nearly reached statistical significance (p<0.1) are also reported.

Results

Patient Experience

Among the 62 patients who received treatment with either the Magstim (N=31) or the NeuroStar (N=31) device, 38 (61%) were female and 24 (39%) were male. All patients completed their treatment course. Forty-four patients (71%) received a 10-Hz left-sided treatment protocol, 11 (18%) received a 20-Hz left-sided protocol, and seven (11%) received a 1-Hz right-sided protocol. The portion of patients receiving 1-Hz, 10-Hz, and 20-Hz protocols on each device was approximately equal with no significant differences. Any patient receiving treatment with the Deep TMS H-Coil System (not included in the analyses) was treated with the standard 18-Hz protocol for this device.
The Magstim and NeuroStar groups did not differ significantly with regard to clinical outcome, with an overall clinical response rate of 48% as defined by a 50% pre-to-post decrease in scores on the Beck Depression Inventory-II and 22.5% full remission rates.
However, patients treated with the NeuroStar device reported having greater confidence that the treatment was being performed correctly compared with those treated with the Magstim device (p<0.05). Conversely, with regard to tolerability, patients treated with the Magstim device reported less pain in the last week (p<0.05) and less pain on average (p<0.05) compared with those treated with the NeuroStar device. Magstim patients also reported feeling more tired on average than their NeuroStar counterparts, although this difference fell short of statistical significance (p=0.063). Cohen’s d was calculated to be 0.40, indicating a medium effect size for this finding.
Among the different protocols, the patients assigned to the 20-Hz protocol reported significantly more anxiety (ANOVA, p<0.05) and endorsed feeling less safe (ANOVA, p<0.05) compared with those assigned to the 1-Hz and 10-Hz protocols.

Technician Experience

Due to the timing of its acquisition, the Berenson Allen Center utilized the Deep TMS H-Coil System device for fewer than 500 sessions versus the >2,500 sessions performed with both the Magstim and NeuroStar devices during the pertinent period of data generation.
On average, over the incongruent sample, the technicians reported feeling that both the Magstim and NeuroStar devices were significantly easier to use than the Deep TMS H-Coil System device (p<0.001). Additionally, they found the former two devices to be more reliable (p<0.001) and better tolerated (p<0.001). Furthermore, the technicians reported greater confidence in using the Magstim and NeuroStar devices than the Deep H-Coil System TMS device (p<0.01) and that they would be more likely to recommend the two former devices to other treatment centers (p<0.001).
Technicians’ reported no significant differences between the Magstim and NeuroStar devices in the aforementioned areas (Table 2).
TABLE 2. Average Technician Ratingsa
VariableNeuroStarMagstimBrainsway Deep TMS H-Coil System
Easy to use4.34.62.6
Well tolerated by patients4.44.42.29
Reliable4.24.42.71
Confident in device efficacy4.74.42.57
Would recommend to other providers4.64.62.14
a
Patient ratings were on a scale from 1 to 5, ranging from “strongly disagree” to “strongly agree.” In each item, the Deep TMS H-Coil System device was significantly different (p<0.05) from the NeuroStar and Magstim devices, which were not significantly different from each other.

Discussion

To our knowledge, this study provides the first evidence-based investigation of patient and technician attitudes toward distinct TMS devices used in the treatment of depression. Each device utilizes distinct approaches to clinical applications of TMS, and each device manufacturer has approached the clinical market in distinct ways. Therefore, it is valuable for the field to better understand patient and technician experiences as they relate to various devices. Our analysis was uncorrected for multiple comparisons and, as such, should be considered exploratory.
While there were notable areas of overlap and functional equivalence, including overall efficacy between at least two of the devices, there were also respective strengths and weaknesses across devices. One meaningful finding was that patients who received treatment using the NeuroStar machine were more confident that the treatment was performed correctly, despite technicians reporting no differences between confidence using the NeuroStar device and the Magstim device. The NeuroStar device notably features a proprietary medical chair and functional headrest that are necessary components of the treatment, while the Magstim features more versatility and is capable of being used as a hand-held device or with any number of company-endorsed accessories (e.g., chair, coil stand). It is difficult to assess whether this finding of greater patient confidence in the NeuroStar device relates to its appearance, its functionality, or actual technician skill, but the resulting perception is nevertheless experienced by patients in a manner that may in fact enhance their clinical experience.
In contrast, with regard to tolerability, patients treated with the Magstim device reported less pain than their counterparts treated with the NeuroStar device. It is worth speculating about a plausible connection between this finding and the previously discussed increased patient confidence in the NeuroStar device, since it is possible that patients’ perception of pain with the NeuroStar device in some way contributed to their confidence that the device was working. Notably, however, none of these perceptions seemingly affected the likelihood of choosing the same device again, since there were no differences between groups on that item.
The data also shed some light on patient experience with distinct frequency protocols used in the treatment of depression. Though no longer a first-line approach for most patients, the 20-Hz protocol has been used extensively in the past at our center and elsewhere and seems to be associated with negative patient experiences of increased anxiety and feelings of being unsafe during the treatment. We did not obtain qualitative data that could illuminate why patients felt this way, and it is worth noting the many factors that might be different between the groups analyzed (including depression duration). However, one could speculate that the intensity of the sound or more rapidly repetitive sensations of discomfort, even in a shorter protocol, may have played a role.
The Brainsway Deep TMS H-Coil System5 device is perhaps the most distinctive as a result of its H-coil design. However, due to the low number of patients enrolled during the inclusion period, the available results could not be used in the patient attitudes analysis. We were able to analyze technician attitudes toward this device because of a 100% response rate among our technicians over the last several years, but these data should be observed with caution because clinical use of this device was highly limited compared with that of the other two devices analyzed. Acknowledging this limitation, our technicians consistently found the Brainsway Deep TMS H-Coil device wanting in the areas of ease of use, tolerability, and reliability efficacy, as well as in terms of likelihood to recommend use of the device to other providers. However, we feel that it would be premature to make meaningful conclusions about these findings because the data obtained featured a sample from only one clinical research center where device familiarity and the center’s academic and research culture may have skewed the results.
We provided the survey to the director of another local TMS center (McLean Hospital, Belmont, Mass.), where the Brainsway Deep TMS H-Coil device has been used extensively alongside the NeuroStar device. No differences in staff experiences were reported, adding validity to the concern that experience with a particular device is highly influential in determining attitudes toward it. Pending greater data collection and expansion to other centers, we recommend that the results be reviewed cautiously.
Other limitations of this study include 1) that it was entirely retrospective in nature and featured small numbers of respondents in both the Brainsway Deep TMS H-Coil device group and the 1-Hz and 20-Hz protocols, respectively, and 2) the amount of time between treatments and survey administration, which varied and was as long as 3 years for some participants.
Although they share similarities, the TMS devices used to treat medication-refractory depression are distinct and confer different patient and technician experiences. TMS providers must take into consideration the existing evidence about device-specific qualities when making decisions about which treatment to offer patients. Furthermore, as increasing numbers of devices and protocols gain popularity, it is important for the field to monitor these parameters in order to optimally match patients with appropriate and individualized treatment.

Footnote

The content of this article is solely the responsibility of the authors and does not necessarily represent the official views of any of the study supporters.

References

1.
Burt T, Lisanby SH, Sackeim HA: Neuropsychiatric applications of transcranial magnetic stimulation: a meta analysis. Int J Neuropsychopharmacol 2002; 5:73–103
2.
George MS, Lisanby SH, Avery D, et al: Daily left prefrontal transcranial magnetic stimulation therapy for major depressive disorder: a sham-controlled randomized trial. Arch Gen Psychiatry 2010; 67:507–516
3.
O’Reardon JP, Solvason HB, Janicak PG, et al: Efficacy and safety of transcranial magnetic stimulation in the acute treatment of major depression: a multisite randomized controlled trial. Biol Psychiatry 2007; 62:1208–1216

Information & Authors

Information

Published In

Go to The Journal of Neuropsychiatry and Clinical Neurosciences
Go to The Journal of Neuropsychiatry and Clinical Neurosciences
The Journal of Neuropsychiatry and Clinical Neurosciences
Pages: 242 - 245
PubMed: 29366375

History

Received: 26 July 2017
Revision received: 18 September 2017
Accepted: 23 September 2017
Published online: 25 January 2018
Published in print: Summer 2018

Keywords

  1. Transcranial Magnetic Stimulation
  2. TMS
  3. Neuromodulation
  4. Noninvasive Brain Stimulation

Authors

Details

Brady Lonergan, M.D.
From the Department of Psychiatry, Harvard Medical School and Beth Israel Deaconess Medical Center, Boston (BL, APS); the Berenson-Allen Center for Noninvasive Brain Stimulation, Division of Cognitive Neurology, Department of Neurology, Harvard Medical School and Beth Israel Deaconess Medical Center, Boston (EN, CL, CH, DZP, AP-L, APS); and McLean Hospital, Belmont, Mass. (OGM).
Eliza Nguyen
From the Department of Psychiatry, Harvard Medical School and Beth Israel Deaconess Medical Center, Boston (BL, APS); the Berenson-Allen Center for Noninvasive Brain Stimulation, Division of Cognitive Neurology, Department of Neurology, Harvard Medical School and Beth Israel Deaconess Medical Center, Boston (EN, CL, CH, DZP, AP-L, APS); and McLean Hospital, Belmont, Mass. (OGM).
Cara Lembo
From the Department of Psychiatry, Harvard Medical School and Beth Israel Deaconess Medical Center, Boston (BL, APS); the Berenson-Allen Center for Noninvasive Brain Stimulation, Division of Cognitive Neurology, Department of Neurology, Harvard Medical School and Beth Israel Deaconess Medical Center, Boston (EN, CL, CH, DZP, AP-L, APS); and McLean Hospital, Belmont, Mass. (OGM).
Carrie Hinchman
From the Department of Psychiatry, Harvard Medical School and Beth Israel Deaconess Medical Center, Boston (BL, APS); the Berenson-Allen Center for Noninvasive Brain Stimulation, Division of Cognitive Neurology, Department of Neurology, Harvard Medical School and Beth Israel Deaconess Medical Center, Boston (EN, CL, CH, DZP, AP-L, APS); and McLean Hospital, Belmont, Mass. (OGM).
Oscar G. Morales, M.D.
From the Department of Psychiatry, Harvard Medical School and Beth Israel Deaconess Medical Center, Boston (BL, APS); the Berenson-Allen Center for Noninvasive Brain Stimulation, Division of Cognitive Neurology, Department of Neurology, Harvard Medical School and Beth Israel Deaconess Medical Center, Boston (EN, CL, CH, DZP, AP-L, APS); and McLean Hospital, Belmont, Mass. (OGM).
Daniel Z. Press, M.D.
From the Department of Psychiatry, Harvard Medical School and Beth Israel Deaconess Medical Center, Boston (BL, APS); the Berenson-Allen Center for Noninvasive Brain Stimulation, Division of Cognitive Neurology, Department of Neurology, Harvard Medical School and Beth Israel Deaconess Medical Center, Boston (EN, CL, CH, DZP, AP-L, APS); and McLean Hospital, Belmont, Mass. (OGM).
Alvaro Pascual-Leone, M.D., Ph.D.
From the Department of Psychiatry, Harvard Medical School and Beth Israel Deaconess Medical Center, Boston (BL, APS); the Berenson-Allen Center for Noninvasive Brain Stimulation, Division of Cognitive Neurology, Department of Neurology, Harvard Medical School and Beth Israel Deaconess Medical Center, Boston (EN, CL, CH, DZP, AP-L, APS); and McLean Hospital, Belmont, Mass. (OGM).
Adam P. Stern, M.D. [email protected]
From the Department of Psychiatry, Harvard Medical School and Beth Israel Deaconess Medical Center, Boston (BL, APS); the Berenson-Allen Center for Noninvasive Brain Stimulation, Division of Cognitive Neurology, Department of Neurology, Harvard Medical School and Beth Israel Deaconess Medical Center, Boston (EN, CL, CH, DZP, AP-L, APS); and McLean Hospital, Belmont, Mass. (OGM).

Notes

Send correspondence to Dr. Stern; email: [email protected]

Competing Interests

Dr. Morales has served as a consultant for Brainsway Ltd. and as Senior Medical Advisor for Achieve TMS East and Achieve TMS. All other authors report no financial relationships with commercial interests.

Funding Information

National Alliance for Research on Schizophrenia and Depression10.13039/100009670: Young Investigator Award
Harvard Catalyst10.13039/100007299: UL1 RR025758, UL1 TR001102
National Institutes of Health10.13039/100000002: (R01HD069776, R21 NS082870, R01NS073601, R21 MH099
Supported in part by the Berenson-Allen Foundation, the Sidney R. Baer Jr. Foundation, NIH (grants R01HD069776, R21 NS082870, R01NS073601, R21 MH099196, R21 NS085491, and R21 HD07616), the Brain and Behavior Research Foundation, and Harvard Catalyst: The Harvard Clinical and Translational Science Center (NCRR and the NCATS NIH, UL1 RR025758, UL1 TR001102), as well as financial contributions from Harvard University and its affiliated academic health care centers.

Metrics & Citations

Metrics

Citations

Export Citations

If you have the appropriate software installed, you can download article citation data to the citation manager of your choice. Simply select your manager software from the list below and click Download.

For more information or tips please see 'Downloading to a citation manager' in the Help menu.

Format
Citation style
Style
Copy to clipboard

View Options

View options

PDF/EPUB

View PDF/EPUB

Login options

Already a subscriber? Access your subscription through your login credentials or your institution for full access to this article.

Personal login Institutional Login Open Athens login
Purchase Options

Purchase this article to access the full text.

PPV Articles - Journal of Neuropsychiatry and Clinical Neurosciences

PPV Articles - Journal of Neuropsychiatry and Clinical Neurosciences

Not a subscriber?

Subscribe Now / Learn More

PsychiatryOnline subscription options offer access to the DSM-5-TR® library, books, journals, CME, and patient resources. This all-in-one virtual library provides psychiatrists and mental health professionals with key resources for diagnosis, treatment, research, and professional development.

Need more help? PsychiatryOnline Customer Service may be reached by emailing [email protected] or by calling 800-368-5777 (in the U.S.) or 703-907-7322 (outside the U.S.).

Media

Figures

Other

Tables

Share

Share

Share article link

Share